Gasboard—3000 flue gas analyzer can be used for measurement of the concentration of SO
2 ,NO ,CO ,CO
2 and O
2 components in the flue gases. It is based on the micro-flow type non-dispersion infrared (NDIR) method for SO
2 ,NO ,CO ,CO
2 and O
2 is measured by electrochemical cell method. This analyzer is designed with 19 inches 3U smaller physical dimensions.
Applications :
The general application is Continuous Emission Monitoring System(CEMS), combustion efficiency monitoring for various industrial furnaces or refuse incinerator and boiler.
Specifications:
Measure Gas
|
Technology
|
Range
|
Resolution
|
Precision
|
RepetitionError
|
SO2
|
Micro-flow NDIR
|
2000ppm
|
1ppm
|
≤1%
|
≤1%
|
NO
|
Micro-flow NDIR
|
2000ppm
|
1ppm
|
≤1%
|
≤1%
|
O2
|
Electrochemical
|
25%
|
0.01%
|
≤3%
|
≤2%
|
Note: Measurement range can be customized by the requirement.
|
Measure Gas
|
Technology
|
Range
|
Resolution
|
Precision
|
RepetitionError
|
CO
|
Micro-flow NDIR
|
5000ppm
|
1ppm
|
≤1%
|
≤1%
|
CO2
|
Micro-flow NDIR
|
25%
|
0.01%
|
≤1%
|
≤1%
|
Note:Measurement range can be customized by the requirement.
|
Gas flow
|
0.7 ~1.2 L/min
|
Pressure of gas inlet
|
<1Bar
|
Sampling gas requirement
|
Remove water vapor,dust(<1um) and oil etc.
|
Response Time(TD+T90)
|
<10s (NDIR)
|
Warm-up Time
|
15min
|
Communication Interface
|
RS232/RS485
|
Output
|
4~20 mA
|
Alarm
|
Alarm
|
Work Temperature
|
0~50℃
|
Relative Humidity
|
5~85%
|
Ambient Pressure
|
86~108kPa
|
Power Source Supply
|
10V-220V±44V AC 50Hz±1Hz
|
Dimension
|
433mm×420mm×132mm(width×length×height)
|
Weight
|
12kg
|